8.2 DET

796 Titroprocessor, Instructions for use

132

132

The various conditions governing the discontinuation of initial dosing can be
combined at will.
N.B.: Initial dosing is discontinued on attainment of any of the start condi-
tions, be it the defined absolute/relative start volume, the defined measured
value or the signal slope. The actual titration will then start immediately after
the pre-set pause.

predef. start volume

meas. value

(reached)

Slope

Vol [mL]

U [mV]

actual initial dosing

The volume actually dosed is filed in the

SV system variable and can be

called up for calculations.

Volume

Definition of the start volume

absolute | relative | off

The defined start volume can be entered as an absolute
volume, or as a relative volume as a function of the sample
size (recommended). Various other fields* will need to be
edited depending on what is entered in this field

(see be-

low).

Size

*

Absolute volume (with abs. start volume)

0…999.99 mL

If an absolute start volume is entered, the start volume
should be kept small enough to ensure that sufficient vol-
ume remains up to the equivalence point for the dynami-
cally controlled titration.

Factor

*

Calculation factor (with rel. start volume)

–99999…0…99999

The actual volume of the start volume to be added will be
calculated as follows:

Volume in mL = sample size

∗∗ factor

.

The maximum start volume that can be added is 1000 mL.
If the calculation results in a higher value, the start volume
is automatically reduced to 1000 mL.