13.3 Key cards and access control
796 Titroprocessor, Instructions for use
274
274
Softkeys
[File]
opens the file dialog window for storing or loading an ac-
cess control file (*.acf)
[Change
passwd.]
opens the dialog window for entering a new password.
[Status
defin.]
opens the dialog window for modifying the status defini-
tions.
[Set all
open]
places all settings for keys and softkeys to 'open'.
13.3.2 Status definitions
The [Status defin.] softkey opens the dialog page where the status
definitions can be modified, see also page 82.
ЪДДД Status definitions ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿
³ ³
³ Status settings for Standard User Level: ³
³ ³
³ Settings edit. Settings edit. ³
³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³
³ ³
³ Sample silo off yes Load methods from Internal ³
³ ³
³ Auto start off no Auto save data off no ³
³ No. of samples 9999 no Save data to Internal no ³
³ ³
³ Statistics off yes Clear working area yes ³
³ No. data records 2 yes Show method window auto ³
³ ³
³ Sample changer no ³
³ ³
³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³
³ ³
³ User yes ³
³ Request user name no ³
³ ³
³ ³
ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄQUITÄÙ
Most of the settings can be declared as editable or not editable.
Sample silo
Switches on or off the sample silo.With sample silo 'on',
the <SAMPLE>key on the keyboard opens the silo in-
stead of the (single) sample input dialog.
edit. yes/no
: with 'no' it is not possible to change
the status of the sample silo in the status window.
Auto start
Switches on or off the auto start function for processing a
sample series.
edit. yes/no
: with 'no' it is not possible to change
the status of the auto start function in the status window.