13.1.12 Setting the servicing interval

After finishing the the test described above, update the service notes. Press
the softkeys [>> >>] and [Service notes] on the Service diagnosis page.

ЪДДД DIAGNOSIS WINDOW ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿¿

³ ³³

³ S E R V I C E D I A G N O S I S ³³

³ -------------------------------- ³³

³ ³³

³ Choose diagnosis program by number or with softkey. ³³

³ Programs can be quit with key <QUIT> or continued by key <Enter>. ³³

³ ³³

³ <f1> Memory and Data card tests ³³
³ ЪДДД

Servicing notes

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿ ³³

³ ³ ³ ³³
³ ³ Instrument ident.

Chem.1

³ ³³

³ ³ Last servicing 1999-12-20 17:12:38 ³hmetic ³³

³ ³ Service engineer C. Daniels ³ ³³

³ ³ ³ ³³

³ ³ Servicing interval 12 month(s) ³ ³³

³ ³ ³ ³³

³ АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ ³³

³ ³³

³ ³³

³ Within arithmetic set float.point error to: error Message ³³

³ ³³

³ ³³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД Quit ДЩЩ

ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿

³Set new³ ³ ³ ³ ³ ³ ³ ³ ³ ³ ³

³date ³ ³ ³ ³ ³ ³ ³ ³ ³ ³ ³

АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ

The Instrument identification should be set once and never be changed later
on.
Enter your name in the field 'Service engineer'.
The servicing interval should be set to a reasonable value. We recommend
to choose 12 to 24 months. After elapsing of the servicing interval, a screen
message will appear after start-up of the instrument. The user will be urged
to call the servic engineer.

Most important. Update the 'last servicing' entry by pressing the [Set new
date] softkey.

Press <QUIT> twice to close the diagnosis section of the 796 Titroproces-
sor.