12.1 Error messages

796 Titroprocessor, Instructions for use

246

246

MEAS/CAL

MEAS/CAL 1:

cycle time too short

The selected measuring interval is shorter than the waiting time of the drift-
controlled individual measurements. This can lead to irregular cycle times
for the series of measurements. If necessary correct the waiting time or
switch off the measurement drift control.

MEAS/CAL 2:

measured value out of buffer tolerance

The measured value of the buffer solution does not correspond with the
nominal buffer value. Check the sensor and buffer solution.

MEAS/CAL 3:

same buffer (?)

The same buffer value has been found twice in succession. The buffer solu-
tion has probably not been changed.

MEAS/CAL 4:

buffer temperature out of tolerance

For a correct calibration all buffer solutions in a series must have approxi-
mately similar temperatures. A temperature has been found which is outside
the tolerated scatter range of 2°C.

MEAS/CAL 5:

invalid measuring value

An invalid measuring value has been found. Check the sensor.

MEAS/CAL 6:

irregular calibration, save anyway?

A fault has occurred during the measurement of a buffer solution. Press
<n> to accept the buffer value anyway, or <a> to interrupt the calibration.

MEAS/CAL 7:

irregular calibration, buffer will be ignored

A fault which cannot be more closely specified has occurred during the
measurement of a buffer solution. The buffer will be ignored for the calibra-
tion.

MEAS.INP.

MEAS.INP. 1:

meas.input 'X' busy, press [Start / Stop]

It is not possible to alter the measuring mode while a measurement is taking
place. The measurement should first be stopped.

MEAS.INP. 2:

meas.input 'X' already locked

A manual measurement cannot be made if the selected measuring input or
sensor is occupied with a running method.

MEAS.INP. 3:

meas.input 'X' not found

The selected measuring input is not built into the instrument.