8.3 MET

796 Titroprocessor, Instructions for use

144

144

8.3 MET

There is a choice of four different

MET modes for controlled titrations with

fixed volume steps. They differ only with regard to the measuring quantity:

MET_PH*

Measuring quantity pH

MET_U*

Measuring quantity mV

MET_UPOL*

Measuring quantity µA (adjustable pol. voltage in mV)

MET_IPOL*

Measuring quantity mV (adjustable pol. current in µA)

The following information applies to all

MET modes. Discrepancies are indi-

cated.

MET_PH*

monotonic equivalence point titration

Reagent Sensor

Concentr. Meas. input A1
Dos. drive A1

applies for

MET_PH*

, MET_U*, MET_UPOL*, MET_IPOL*

MET modes create the same system variables as DET modes, see
page 129.

Assignment parameters of the MET modes:

The assignment parameters (comment, reagent … etc.) of the

MET modes

coincide with those of the

DET modes, see page 130f.

8.3.1 [ * Param ] Specific parameters for MET modes

The specific parameters of the corresponding

MET mode will be found un-

der the [* Param ]softkey.

ЪДДД

MET_PH* Parameters

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ Start conditions ³

³ Volume off Meas.val. pH off ³

³ Slope off pH/mL ³

³ Dos. rate max mL/min Pause 0 s ³

³ ³

³ Titration conditions ³

³ Adjustment medium Time recording on ³

³ Temperature 25.0 øC ³

³ ³

³ Stop conditions ³

³ Volume absolute Meas.val. pH off ³

³ Size 99.99 mL EP# 9 ³

³ fill. rate max mL/min Stop time off s ³

³ Mpt. overflow on ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ