12. . Troubleshooting

Troubleshooting

796 Titroprocessor, Instructions for use

239

239

DIAGNOSIS

If error messages occur during diagnosis please contact the service de-
partment of your local Metrohm agency.

DIALOG

DIALOG 1:

method name changed, unable to store field cont.

A new method has been loaded and a determination started in the back-
ground during editing. This means that the edited field content cannot be
stored with the previous determination.

DIALOG 2:

method load disabled: quit method editor

The Titroprocessor cannot load a method automatically as defined in the
sample silo, as alterations to the current method are being made in the
method editor at the same time. Close the method editor.

DIALOG 3 + 4:

RS232 interface busy

An attempt has been made to access a RS232 interface while this is occu-
pied with a different task. Wait until the interface is free.

DIALOG 6:

command list full

The maximum number of commands in a method has been reached. Delete
unnecessary or empty command lines.

DIALOG 7:

method load impossible while editing

The Titroprocessor cannot load a method automatically if entry fields rele-
vant to the current determination are being edited at the same time. Stop
editing or interrupt the method run with the <HOLD> key.

DIALOG 8:

abort <a> restarts the Titroprocessor

The memory organization has been damaged by unexpected actions in
such a way that the Titroprocessor must be restarted. The restart is carried
out automatically by pressing the <a> key.

DIALOG 9:

no measured values available

As no measurements are present for the current determination it is not pos-
sible to produce a measurement curve.

DIALOG 12:

"EMPTY" not possible for 685 Dosimat "XX"

The EMPTY function is not available for a 685 Dosimat.

DIALOG 13:

tubing configuration of Dosimat "XX" not found

The tubing set with the tubing parameters given for dosing drive XX is not
present. It may have been deleted. Enter a different tubing set or create a
new definition under [Config] - [Tubings].