12.1 Error messages

796 Titroprocessor, Instructions for use

234

234

CHANGER 17:

illeg. rack and swing head combination, changer 'X'

A swing head is required on the sample changer for the rack placed

on it. This must be entered in the sample changer configuration, see

[Config.] [Sample changer] [Changer 1(2)].

A swing head has been entered in the sample changer configuration.

Use a sample rack designed for it or switch off the swing head (see

above).

CHANGER 21:

changer X busy

An attempt has been made to access changer X, although this is either

already occupied with a different task or is not present at all.

CHANGER 22:

changer X unexpected error

A sample changer fault has occurred which cannot be specified more

closely. have you entered a sample changer address (see

<STATUS>key)? check the auto start settings. Try to remedy the error by

switching the instrument off and on again.

CHANGER 23:

changer 'X' special beaker not defined

A particular special beaker is required for the method used. Define the

appropriate special beaker for the sample rack used (softkeys [Config.]

[Sample changer] [Special beaker]) or alter the method (command

Changer).

CHANGER 24:

changer 'X' after MOVE_B no beaker present

The required special beaker has not been found. The sample changer

can be manipulated manually in the HOLD condition. Place the special

beaker in the corresponding position and press <HOLD/CONT>.

CHANGER 25:

changer 'X' wrong rack

The method used requires a particular sample rack which apparently is

currently not placed on the sample changer. The sample changer can be

manipulated manually in the HOLD condition. Place the correct rack on

the changer and press <HOLD/CONT>.

CHANGER 27:

changer 'X'; missing sample beaker

The next sample beaker has not been found. The sample changer can be

manipulated manually in the HOLD condition. Place the correct sample

beaker in the appropriate position and press <HOLD/CONT>.

CHANGER 28:

changer 'X'; Pump missing or not configured

The selected pump is not present or has not been configured. Check the

configuration under [Config.] [Sample changer] [Changer 1(2)].