8. Command reference

796 Titroprocessor, Instructions for use

159

159

8.6 KFT

Two different

KFT modes are available for titrimetric water determination

with Karl Fischer reagents; they differ only in the measuring method:

KFT_UPOL*

measured quantity µA (selectable pol.voltage in mV)

KFT_IPOL*

measured quantity mV (selectable pol.current in µA)

The following information applies to both

KFT modes. Any variations are

noted.

KFT_IPOL*

Karl Fischer titration

Reagent Sensor

Concentr. Meas. input Apol
Dos. drive A1

applies to

KFT_IPOL*, KFT_UPOL*

KFT modes produce the same system variables as DET modes, see page
129. In addition the following variables are provided for calculating a drift
volume reduction (see page 155):

DOD

Dosing duration during the titration, cumulative

ITD

Time duration between conditioning and titration

Assignment parameters for KFT modes:

The assignment parameters (Comment, Reagent, etc.) for the

KFT modes

are the same as those for the

DET modes, see page 129f.

8.6.1 Specific parameters for KFT modes [ * Param ]

The specific parameters for the respective

KFT modes can be found under

the softkey [* Param ].

ЪДДД

KFT_IPOL* Parameters

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ Start conditions ³

³ Volume off Pause 1 0 s ³

³ Pause 2 0 s ³

³ Extract.time 0 s ³

³ ³

³ Titration conditions ³

³ Direction neg. Temperature 25.0 øC ³

³ End-point 250 mV Time recording on ³

³ Adjustment medium Derivation off ³

³ Meas.interval off s ³

³ I(pol) 50 µA ³

³ Stop conditions ³

³ Volume absolute Stop time off s ³

³ Size 99.99 mL Mpt. overflow off ³

³ Fill. rate max mL/min ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ