12.1 Error messages

796 Titroprocessor, Instructions for use

254

254

STORAGE 10:

terminate "KeyCTRL mode" first

For the required action the key simulation mode 'keyCTRL' of the RS232 in-
terface must be switched off. This can be carried out under [Config.] [Inter-
faces].

STORAGE 11:

"RS 232 Interface" still active

There is still a connection to an external instrument via the RS232 interface.
This data communication should first be terminated.

STORAGE 12:

invalid file name

The file name is not in accordance with the conventions for the naming of
files, see page 74.

STORAGE 13:

RS232 interface, check settings

The interface parameters do not meet the requirements of the selected ac-
tion. Correct the settings under [Config.] [Interfaces].

STORAGE 14:

copy process aborted

A copying process could not be completed owing to an operator action. The
selected file(s) have not been copied.

STORAGE 15:

unexpected file block size

STORAGE 16:

memory block size violation

Faults have occurred during file transfer. Check the interface parameters of
the PC software.

STORAGE 17:

undefined error

A fault has occurred which cannot be specified more closely. Please con-
tact the service department of your local Metrohm agency.

STORAGE 18:

invalid type of file

The selected file cannot be stored in the given memory area as it does not
meet the specified type.

STORAGE 19:

invalid file

The selected file is damaged or of an unsuitable type.

STORAGE 20:

remove balance from interface X

No balance can be active on the same interface during file transfer. Switch it
off or connect it to a different interface.