3.4 Buret units

796 Titroprocessor, Instructions for use

50

50

Dosing and filling speeds for PREP and EMPTY

ЪДДД

Tubings

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ Dosino 700: Dosing units, rates ³

³ ³

³ ³

³ ³

³ # Tye Dosiing rate Filling rate ³

³ ДД ДДДДДДДД ДДДДДДДДДДДДД ДДДДДДДДДДДДД ³

³ 1 Default max mL/min max mL/min ³

³ 2 Changer max mL/min max mL/min ³

³ 3 mL/min mL/min ³

³ 4 mL/min mL/min ³

³ 5 mL/min mL/min ³

³ 6 mL/min mL/min ³

³ 7 mL/min mL/min ³

³ 8 mL/min mL/min ³

³ 9 mL/min mL/min ³

³ 10 mL/min mL/min ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ

The dosing and filling speeds for the

PREP and EMPTY functions of the

dosing units (

EMPTY only for 700 Dosino) are defined here. They only

apply to these functions.

Type

Name of the buret unit, see above.

Dosing

rate

The maximum value for the dosing speed depends on
the cylinder size of the buret used.

Buret cylinder

Max. speed

1 mL

3 mL/min

2 mL

6 mL/min

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

For smaller buret cylinders a value of up to 150 mL/min
can also be entered, but when the function is carried out
the dosing speed will automatically be reduced to the
largest possible value. Whenever possible '

max' should

be selected, except for viscous reagent solutions.

Filling

rate

The same applies for the filling speed. Please note that with
viscous solutions too rapid filling can cause a partial
vacuum in the cylinder (bubble formation!).