Metrohm 796 Titroprocessor User Manual

Page 255

Advertising
background image

12

12. . Troubleshooting

Troubleshooting

796 Titroprocessor, Instructions for use

249

249

MONITORING 4:

auto save aborted; XXX bytes needed

There is too little space available in the target memory. Delete files which are
no longer required.

MONITORING 7:

new method while statistics ON

A new method has been loaded or the existing method has been altered
while statistics is switched on. During the statistical calculation this could
lead to results with the same name which have been produced in different
ways being calculated with each other.

MONITORING 8:

working memory full; delete determinations

There is not enough working memory space available for the next determi-
nation. Delete determinations from the working memory (in the file man-
ager). Also reduce the set value for the statistics counter, if this is neces-
sary. This value lays down how many determinations are to be held in the
working memory before the working memory is cleared (during the next de-
termination).

MONITORING 10: method autoload disabled: quit method editor

While alterations are being made in the method editor it is not possible to
load a method automatically, as can be defined for a sample series in the
sample silo. Close the method window with <QUIT> in order to allow the
sample series to proceed.

MONITORING 11: GLP data not saved, irregular termination

A GLP validation has not been concluded properly.

MONITORING 12: check GLP data

The validation interval for a validation method has expired. Carry out a new
validation.

MONITORING 13: Do you want to reset all statistics data?

Query before resetting the statistics. The working memory will be cleared.

MONITORING 14: method line XX : no condition!

The method contains a CASE or EXIT command without any condition.
Check the method line 'xx'.

MONITORING 15: Impossible to execute "SHOW" while editing

Because of a SHOW command in the current method a particular dialog
page is to be shown; this is not possible while an entry field is being edited.
Close the editing mode. The required page will be shown after the next
SHOW command.

Advertising