11.4 File backup

796 Titroprocessor, Instructions for use

230

230

If problems occur during data reception it may be necessary to
deactivate the FIFO buffer of the RS232 interface in the system settings
of the computer or to reduce the baud rate of both instruments to 2400
Baud. Consult the instructions for use of the Metrodata 796 backup
software.

In order to start data transmission the

‘Receive’ or ‘Send’ mode of the

backup software at the PC must be switched on before the [Send to RS]
or [Receive from RS] softkey of the Titroprocessor is pressed. After
selection of the RS232 interface transmission takes place.

Files (method, silo data, determination files) can also be transmitted or
received directly from the file dialog window of the corresponding dialog
page. Select 'Storage' 'RS232 x' and press the [Load] or [Save]
softkey.