13.1 Diagnosis

796 Titroprocessor, Instructions for use

266

266

13.1.9 To check Analog Interface A

With this diagnostic step the analog inputs of Interface A can be checked for their function-
ality. For this test the testbox with article number 1.767.0010 is required.

1.

Prepare instrument for diagnostic test (see chap. 13.1.2).

2.

<Analog interf.>

Ä

On the screen there appears a window with the choice of selecting between analog interface A or B.

3.

With the cursor key, mark the letter A to select analog interface A.

4.

<ENTER>

5.

Connect Simulator to Titroprocessor accord. to Fig. 2.

6.

<All inp measure>

Ä

The measurements are continuously updated and presented as follows::

Test date: ??-??-?? ??:??:??

Analog input measuring test

!!!!!!!!!!!!!!!!!!!!!!!!!!!

!!!!!!!!!!!!!!!!!!!!!!!!!!!

Setting polarizing to: U=300 mV, I=10 uA

Display values as average of 6 meas. cycles

Analog interface A, Calib date: ??-??-?? ??:??:??

!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!

!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!!

Input 1 : ?x??????? ?.???? mV ?.???? mV/min ?.?? °C

Input 2 : ?x??????? ?.???? mV ?.???? mV/min ?.?? °C

Input Diff : ?x??????? ?.???? mV ?.???? mV/min ?.?? °C

Input Ipol : ?x??????? ?.???? mV ?.???? mV/min ?.?? °C

Input Upol : ?x??????? ?.???? mV ?.???? mV/min ?.?? °C

Fig. 1

7.

Check the following measuring values:

column C (see Fig. 1)

column E (see Fig. 1)

Input 1 :

+U / direct

±

0.4 mV

Pt100

±

0.3°C

Input 2 :

+U / 1G

±

0.4 mV

Input Diff :

0 mV

±

4.0 mV

8.

Connect Simulator to Titroprocessor accord. to Fig. 3.

column C (see Fig. 1)

Input 1 :

+U / 1G

±

0.4 mV

Input 2 :

+U / direct

±

0.4 mV

Input Diff :

0 mV

±

4.0 mV

DIAGNOSIS WINDOW