13.2 Common variables

796 Titroprocessor, Instructions for use

270

270

13.2 Common variables

Common variables can be used to store numerical values permanently in
the Titroprocessor 796. The values of these variables are available at any
time for later determinations. They can be used in calculation formulas or in
a method with the

CASE command as the system variables C1…C19.

The common variables table can be found on the main page by pressing
the [Common vars] softkey. These are always the current (global) com-
mon variables.

ЪДДД Common Variables : actual ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ C# Value Unit Comment Date Time Method Flag ³

³ ДД ДДДДДДДДД ДДДДДД ДДДДДДДДДДДД ДДДДДДДД ДДДДДДДД ДДДДДДДД ДДДД ³

³ 1 1.0014 Titer HCl 1999-12-18 08:47:58 HCl oo ³

³ 2 0.9935 Titer NaOH 1999-12-11 09:35:12 NaOH mo ³

³ 3 ....-..-.. ..:..:.. ³

³ 4 ....-..-.. ..:..:.. ³

³ 5 ....-..-.. ..:..:.. ³

³ 6 ....-..-.. ..:..:.. ³

³ 7 ....-..-.. ..:..:.. ³

³ 8 ....-..-.. ..:..:.. ³

³ 9 ....-..-.. ..:..:.. ³

³ 10 ....-..-.. ..:..:.. ³

³ 11 36.46 g/mol HCl 1999-10-12 10:16:55 by edit nn ³

³ 12 40.00 g/mol NaOH 1999-10-12 10:15:01 by edit nn ³

³ 13 ....-..-.. ..:..:.. ³

³ 14 ....-..-.. ..:..:.. ³

³ 15 ....-..-.. ..:..:.. ³

³ 16 ....-..-.. ..:..:.. ³

³ 17 ....-..-.. ..:..:.. ³

³ 18 ....-..-.. ..:..:.. ³

³ 19 ....-..-.. ..:..:.. ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ

ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿

³ ³ ³ ³ ³ ³ ³ ³ ³ ³Delete ³Delete ³

³ ³ ³ ³ ³ ³ ³ ³ ³ ³ComnVar³ all ³

АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ

Common variables can be entered manually; they can also be produced
automatically during a determination. This is done by defining a calculation
formula which contains a corresponding allocation, e.g. C1= … instead of
R1=…

Description of the columns

C#

The description of the common variable (

C1..C19)

Value

The numerical value of the common variable

Unit

Unit of the common variable. This can be selected from
an existing list with the space bar or the [Select] soft-
key.

Comment

Text field for describing the contents of this variable.

Date / Time

Date and time of creation or last alteration.

Method

Shows the way in which the variable was produced. The
different possibilities:

Advertising