7. Titration and measuring modes

796 Titroprocessor, Instructions for use

107

107

7.4 SET

Titration to a preset endpoint

Application:
SET
modes are rapid titration modes for titrations to a preset endpoint. They
correspond to manual titrations and are suitable for rapid determinations of
non-critical samples.
Their typical feature is controlled reagent dosing depending on the control
difference (actual value – endpoint) with the aim of reaching the preset value
defined as the endpoint as quickly and accurately as possible. Overshoot-
ing is virtually eliminated. No automatic endpoint recognition is carried out.

Important: the endpoint should be selected so that it is near a point of in-
flection of the expected titration curve; it should not lie in an area where the
slope is too steep.
Accurate results in the

SET_PH* mode can only be expected when the

sensor used has been previously calibrated correctly.

U[mV]

V[ml]

t[s]

EP

V[ml]

Initial

addition

Continuous

addition

Control

range

SET: endpoint titration

Reagent addition:

controlled addition,
depending on the control
difference
(actual value – endpoint)

Switch-off criterion:

drift-controlled and/or after
waiting period

7.4.1 Measuring mode

An endpoint titration can be carried out with four different types of meas-
urement:

potentiometric pH measurement (

SET_PH*)

potentiometric voltage measurement (

SET_U*)

amperometric measurement with selectable polarization voltage (

SET_UPOL*)

voltametric measurement with selectable polarization current
(

SET_IPOL*)

7.4.2 Reagent addition

Reagent addition takes place in three different phases. The initial addition
takes place with increasing dosing speed. It begins with the defined 'Min.
Rate' and increases up to the 'Max. Rate'.
Dosing is carried out at the defined 'Max. Rate' until the control range is en-
tered.