8. Command reference

796 Titroprocessor, Instructions for use

211

211

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

SHOW

Page main

Page

Show dialog page

curve | result |stat.| main

Depending on the selected dialog page the switch is ei-
ther made immediately or at the end of the determination.

curve

immediate switch to live curve
display

result

switches to results display at end of
determination

stat.

switches to statistics results display at
end of sample series

main

display immediate switches to
main page

NOOP

No operation, place holder command

The

NOOP command can be used to arrange the command blocks. It does

not trigger any actions in the method run.

BEEP

Acoustic signal

The

BEEP command is used to produce an acoustic signal.

END

End marker

Each method must contain an

END command. This is used to mark the end

of the method. Normally the END command is set in the last method line.
For test purposes in a complex method the insertion of further

END com-

mands may be useful, e.g. in a

CASE sequence. The method will then be

processed until the first

END marker is reached.

STIR_I

Control of internal stirrer connections

The stirrer connections of measuring groups A and B can be switched di-
rectly in a method. Stirrer models 728 (magnetic stirrer), 722 (rod stirrer) or
the 703 or 727 Titration stands (with magnetic stirrer) are suitable for use on
the internal connections of the measuring interfaces.

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

STIR_I

Meas. interface A State on

After switch-on the stirrer connections are also switched on. The stirrers
connected to the measuring groups can also be operated manually; see
<MAN CONTROL> key.