2 hello command – INFICON Cygnus Thin Film Deposition Controller User Manual
Page 112
5 - 8
IP
N 07
4-
37
9-
P1
K
Cygnus Operating Manual
5.4.2 HELLO Command
The format is:
H<Command ID>
Response:
ACK<<Byte>|<Integer>|<String>(<Byte>|<Integer>|<String>...<Byte>|<Integer>|<String>)
>|<Response Error Code>
Structure Version . . . . . . . . . . . . . . . The structure version is increased when new
parameters are added.
Compatibility Version . . . . . . . . . . . . A compatibility version change means that
the internal parameter structure has changed
so dramatically that parameter sets with
different compatibility versions cannot be
used.
Range Version . . . . . . . . . . . . . . . . . A range version change means one or more
parameters have different ranges from
previous versions.
Command ID
Meaning
Response
1
ASCII Name and Version
<String> = "Cygnus Version x.xx"
2
Structure Version,
Compatibility Version,
and Range Version
<Integer><Integer><Integer>
3
Firmware version number
<Float>