3 general parameters, 1 query general parameters, 2 update general parameters – INFICON Cygnus Thin Film Deposition Controller User Manual

Page 120

Advertising
background image

5 - 16

IP

N 07

4-

37

9-

P1

K

Cygnus Operating Manual

5.4.3.3 General Parameters

5.4.3.3.1 Query General Parameters

The message format is:

QG<Command ID>

Command ID = <Byte> (See

Table 5-5

.)

Command Response:

ACK<Integer> (All the QG responses are integers.)|<Response Error Code>

Special Case: QG0 will return all General Parameters, in numeric order.

5.4.3.3.2 Update General Parameters

Update: The message format is:

UG<Command ID><Parameter Value>

Command ID = <byte> See "Command ID" column in table below.

Parameter Value = <integer>

Description: type of parameter value depends on Command ID, see
"units" column of

Table 5-5

.

Command Response:

ACK|<Response Error Code>

NOTE: General parameters may only be updated if all channels are in Ready or

Stop and are not crystal switching on any channel.

Special Case: UG0 will send all General parameters in numeric order.

Table 5-5 General Parameters

QG

Command

ID

Name

Units / Allowed

Values / Notes

Low Limit

High Limit

1

(0x01)

Test On

1 = Yes, 0 = No

0

1

2

(0x02)

Time

Compressed

1 = Yes, 0 = No

0

1

3

(0x03)

Audio

Feedback

1 = Yes, 0 = No

0

1

4

(0x04)

(Not Used)

Int

0

0

Advertising