IP

N 07

4-

37

9-

P1

K

Cygnus Operating Manual

8.4 Crystal Sensor Emulator

IPN 760-601-G1 or 760-601-G2

NOTE: 760-601-G2 is fully compatible with all Thin Film Deposition Controllers.

Crystal Sensor Emulator 760-601-G1 (obsolete) is not compatible for use
with an IC/5, IC/4 or Cygnus.

The Crystal Sensor Emulator option is used in conjunction with the Thin Film
Deposition Controller to rapidly diagnose problems with the Deposition Controller's
measurement system. See

Figure 8-5

.

Figure 8-5 Crystal Sensor Emulator

The Crystal Sensor Emulator may be attached at various points in the
measurement system, from the oscillator to the sensor head. It provides a known
"good" monitor crystal with known "good" electrical connections. Using the
emulator and the controller in a systematic manner provides a fast means of
isolating measurement system, cable, or sensor problems. See

Figure 8-6

.

Figure 8-6 Crystal Sensor Emulator Attachment Points

CAUTION

This product is designed as a diagnostic tool, and is not
intended for use in vacuum. Do not leave the Crystal
Sensor Emulator installed in the vacuum system during
processing.

Female

BNC

Connector

Female

Microdot

Connector

Sensor

Cover

Connector

Thin Film

Controller

Crystal Interface

Unit (Oscillator)

Sensor Head

A

B

C

Advertising