3 rs-232c communications – INFICON Cygnus Thin Film Deposition Controller User Manual

Page 155

Advertising
background image

6 - 13

IP

N 07

4-

37

9-

P1

K

Cygnus Operating Manual

6.3.4.3 RS-232C Communications

RS-232C serial communications is included in the controller as standard
equipment. It is used to remotely control or monitor the Cygnus. An industry
standard 9-pin D-Sub connector is required for the host computer side connection.
Depending on the computer source, all connections may not be necessary. The
length of the cable is limited to fifty feet according to published standards. The
controller interface operates as DTE (Data Terminal Equipment). Pin assignments
are for the Cygnus connector.

Table 6-4 RS-232C Pin Connections

Signal Name

Pin

EIA Name

TX

Transmit Data

2

BA

RX

Receive Data

3

BB

RTS

Request To Send

8

CA

CTS

Clear To Send

7

CB

DSR

Data Set ready

4

CC

SG

Signal Ground

5

AB

(Not Used)

1

DTR

Data Terminal Ready

6

CD

GND

Shield Ground

9