2 update all parameters, 2 channel parameters, 1 query channel parameters – INFICON Cygnus Thin Film Deposition Controller User Manual

Page 114

Advertising
background image

5 - 10

IP

N 07

4-

37

9-

P1

K

Cygnus Operating Manual

5.4.3.1.2 Update All Parameters

Update: The message format is:

UA <All configuration parameters>

Length = 2 bytes (low byte / high byte)
All Channel 1 parameters, all Channel 2 parameters,...all Channel 6
parameters.
All General parameters
All Logic Statements
All Output Types
All Output Names
All Input Names

Command Response:

ACK|<Response Error Code>

NOTE: All channels in the instrument need to be in Ready or Stop for this

command to be accepted.

5.4.3.2 Channel Parameters

5.4.3.2.1 Query Channel Parameters

Query: This command returns a specific parameter or all parameters for a specified
channel.

The message format is:

QC<Command ID><Channel Number>

Command ID = <Byte> See

Table 5-4

.

Channel Number = <Byte> 1 to 6

Command Response:

ACK<<Integer>|<Float>>|<Response Error Code>

Description: Type of response depends on Command ID, see "Data Type" column
of the table below.

Special Case: QC0<Channel Number> will return all Channel parameters for the
given channel, in numerical order.