2 update all parameters, 2 channel parameters, 1 query channel parameters – INFICON Cygnus Thin Film Deposition Controller User Manual
Page 114
5 - 10
IP
N 07
4-
37
9-
P1
K
Cygnus Operating Manual
5.4.3.1.2 Update All Parameters
Update: The message format is:
UA <All configuration parameters>
Length = 2 bytes (low byte / high byte)
All Channel 1 parameters, all Channel 2 parameters,...all Channel 6
parameters.
All General parameters
All Logic Statements
All Output Types
All Output Names
All Input Names
Command Response:
ACK|<Response Error Code>
NOTE: All channels in the instrument need to be in Ready or Stop for this
command to be accepted.
5.4.3.2 Channel Parameters
5.4.3.2.1 Query Channel Parameters
Query: This command returns a specific parameter or all parameters for a specified
channel.
The message format is:
QC<Command ID><Channel Number>
Command ID = <Byte> See
.
Channel Number = <Byte> 1 to 6
Command Response:
ACK<<Integer>|<Float>>|<Response Error Code>
Description: Type of response depends on Command ID, see "Data Type" column
of the table below.
Special Case: QC0<Channel Number> will return all Channel parameters for the
given channel, in numerical order.