7 type of output parameters, 1 query output type, 2 update output type – INFICON Cygnus Thin Film Deposition Controller User Manual
Page 128: 4 status commands

5 - 24
IP
N 07
4-
37
9-
P1
K
Cygnus Operating Manual
5.4.3.7 Type of Output Parameters
5.4.3.7.1 Query Output Type
Query: The message format is:
QT<Output Number>
Output Number = <byte> 0 to 38 (0 = All Outputs)
Command Response:
If Output Number was 1 to 38:
ACK<Output Type>|<Response Error Code>
If Output Number was 0:
ACK<Output 1 Type><Output 2 Type>...<Output 38 Type>|<Response Error
Code>
Output Type = <byte> 0 or 1 (0 = NO, 1 = NC)
5.4.3.7.2 Update Output Type
Update: This command changes the type code for a specified output. Type code
values are 0 or 1, where 0 is Normally Open and 1 is Normally Closed.
The message format is:
UT<Output Number><Output Type>|<Output Number = 0><Output 1
Type><Output 2 Type>...<Output 38 Type>
Output Number = <byte> 0 to 38 (0 = All Outputs)
Output Type = <byte> 0 or 1 (0=NO, 1 = NC)
Response:
ACK|<Response Error Code>
NOTE: Output Types may only be updated if all channels are in Ready or Stop and
are not crystal switching on any channel.
5.4.4 STATUS Commands
Status commands return pertinent information based on a specific request made.
Commands are provided to determine global information which is system level or
channel information. A status code is required for each command. The command
variations include:
SC . . . . . . . . . . . . . . . . . . . . . . . . . . Status of a Channel
SG . . . . . . . . . . . . . . . . . . . . . . . . . . Status of system or instrument level
condition.