4 input name parameters, 1 query input name, 2 update input names – INFICON Cygnus Thin Film Deposition Controller User Manual
Page 122
5 - 18
IP
N 07
4-
37
9-
P1
K
Cygnus Operating Manual
5.4.3.4 Input Name Parameters
5.4.3.4.1 Query Input Name
Query: The message format is:
QI<Input Number>
Input Number = <byte> 0 to 28 (0 = all inputs)
Command Response:
If Input Number was 1 to 28:
ACK<Input Name>|<Response Error Code>
If Input Number was 0:
ACK<Input 1 Name><Input 2 Name>...<Input 28 Name>|<Response Error
Code>
Input Name = <string> Up to 10 characters, null terminated.
5.4.3.4.2 Update Input Names
Update: The message format is:
UI<Input Number><Input Name>|<Input Number = 0><Input 1 Name><Input 2
Name>...<Input 28 Name>
Input Number = <byte> 0 to 28 (0 = all inputs)
Input Name = <string> Up to 10 Characters, null terminated.
Command Response:
ACK|<Response Error Code>
NOTE: Input Names may only be updated if all channels are in Ready or Stop and
are not crystal switching on any channel.