5.4.3.4 Input Name Parameters

5.4.3.4.1 Query Input Name

Query: The message format is:

QI<Input Number>

Input Number = <byte> 0 to 28 (0 = all inputs)

Command Response:

If Input Number was 1 to 28:

ACK<Input Name>|<Response Error Code>

If Input Number was 0:

ACK<Input 1 Name><Input 2 Name>...<Input 28 Name>|<Response Error
Code>

Input Name = <string> Up to 10 characters, null terminated.

5.4.3.4.2 Update Input Names

Update: The message format is:

UI<Input Number><Input Name>|<Input Number = 0><Input 1 Name><Input 2
Name>...<Input 28 Name>

Input Number = <byte> 0 to 28 (0 = all inputs)

Input Name = <string> Up to 10 Characters, null terminated.

Command Response:

ACK|<Response Error Code>

NOTE: Input Names may only be updated if all channels are in Ready or Stop and

are not crystal switching on any channel.