5.4.5 REMOTE Commands

Remote commands perform an action based on the specific command given.
Commands are provided to affect global features which are system level oriented
(RG commands) or channel oriented (RC commands). Every command requires a
remote code and may require a value. Command variations include:

RC . . . . . . . . . . . . . . . . . . . . . . . . . . Affects conditions for a specified Channel.

RG . . . . . . . . . . . . . . . . . . . . . . . . . . Affects system or instrument level conditions.

11

(0x0B)

System
Configuration

2 bytes Configuration Status
0 0x0001 Measurement Board 1 Detected
1 0x0002 Measurement Board 2 Detected
2 0x0004 Measurement Board 3 Detected
3 0x0008 Not Used
4 0x0010 I/O Board 1 Installed
5 0x0020 I/O Board 2 Installed
6 0x0040 I/O Board 3 Installed
7 0x0080 Reserved
8 0x0100 IEEE-488 Board Installed
9 0x0200 Reserved
10 0x0400 Reserved
11 0x0800 Reserved
12 0x1000 Reserved
13 0x2000 Reserved
14 0x4000 Reserved
15 0x8000 Reserved

12

(0x0C)

Timer Status

<Byte>
Timer
Number

0 = Timer Inactive
1 = Timer Active

13

(0x0D)

Timer Value

<Byte>
Timer
Number

4 Bytes, Integer number in tenths of a second

14

(0x0e)

Counter Value

<Byte>
Counter
Number

2 Bytes, Integer number of count

Table 5-9 Status General Commands (continued)