2 - 4

IP

N 07

4-

37

9-

P1

K

Cygnus Operating Manual

8 8 Relay x 14 Output I/O Card (optional)

Provides pin connection for 8 relays rated for 30 V(dc) or 30 V(ac) RMS or
42 V(peak) maximum, and 14 open collector type outputs.

9 6-Channel DAC (standard)

Provides source control voltage or recorder output for 6 channels (BNC
connectors). Outputs are programmable for Source Control or Recorder
function.

10 24-Volt Supply (standard)

Provides one 24 V(dc) supply rated at 1.75 Amps.

11 Fan Outlet

Exhaust opening for the unit’s miniature fan. A grill is attached for safety.

12 AC Power Inlet

Provides a common connector for various international plug sets. The unit is
factory set for 100 V(ac), 120 V(ac), 230 V(ac) or 240 V(ac) service.

13 RS-232C Remote Communication Connector (standard)

Provides a 9-pin RS-232C communications port.

2.3 Displays

Cygnus has many display screens for monitoring and programming processes. At
the top of the hierarchy is the ALL CHANNELS display. The ALL CHANNELS
display allows the user to simultaneously monitor process information for up to six
channels. Three main types of displays, GENERAL PARMS; GENERAL
MAINTENANCE and CHANNEL are accessible from the ALL CHANNELS
DISPLAY: To move from one display to another, use the function keys to the right
of the screen.

Figure 2-4

provides an overview of the Cygnus display hierarchy.