2 - 19
IP
N 07
4-
37
9-
P1
K
Cygnus Operating Manual
SENSOR MAINTENANCE Display Function Keys
F1 CRYSTAL SWITCH
To initiate a crystal switch, press function key F1.
F2 ROTATE XTAL HEAD
If the channel sensor type is a CrystalSix, Crystal12, or Rotary, press function
key F2 to sequentially rotate the sensor head through all positions. This is
useful to initialize a CrystalSix, Crystal12 or Rotary sensor after replacing failed
crystals, or changing Sensor Type parameter.
F3 CLEAR FAILED CRYSTALS
If the channel sensor type is a CrystalTwo, CrystalSix or Crystal12, press
function key F3 to clear all non current failed crystal position numbers
(CrystalSix, Crystal12) or to increment the number of good crystals
(CrystalTwo), if applicable.
F4 CLEAR QUALITY / STABILITY
To clear the Quality and Stability counters, press function key F4.
F5 SOURCE MAINTENANCE
To access the channel source maintenance display press function key F5.
F6 CHANNEL
To return to the Operate display, press function key F6.