5.4.6 Negative Response Error Codes

An error code in the form of a single ASCII character is returned when a command
results in an error. There are two types of Error codes: a Packet Error Code or a
Response Error Code. Both are described under section

section 5.3.2.2,

Response Packet (Instrument to Host Message), on page 5-6

.

5.5 Sample Cygnus Commands

The following illustrate examples of typical message streams.

5.5.1 Remote Communications Example — Updating a

Channel’s Rate Parameter

An example of setting the Rate parameter to 10.00 A/s in Channel 3 by Remote
Communication is shown below.

The Rate parameter is changed by a UC command, since it will Update the
parameter value of a Channel.

The Rate parameter is a "Float" data type, and is a 32 bit number.

The Character "U" = 0x55 (hexadecimal 8 bit value)

The Character "C" = 0x43 (hexadecimal 8 bit value)

Rate Command ID is 20 = 0x14 (hexadecimal 8 bit value)

Channel 3 = 0x03 (hexadecimal 8 bit value)

10 (decimal) = 00002041 (hexadecimal floating point 32 bit value in LSB to MSB
order)

Putting them all together builds the 8 byte message shown as a byte message in
hexadecimal:

5543140300002041

Calculate the message checksum = {(sum of all bytes in the message) modulus
0x0100}

Checksum = {0x55 + 0x43 + 0x14 + 0x03 + 0x00+ 0x00 + 0x20 + 0x41) modulus
0x0100}

Checksum = {(0x0110) modulus 0x100}

Checksum = 0x10

Now to send the message, put a two byte length (LSB, MSB) ahead of the message
and a one byte checksum after the message.

0800554314030000204110