3 determining tooling, Section 7.3, Density g cm – INFICON Cygnus Thin Film Deposition Controller User Manual

Page 160: Tooling (%) tf

Advertising
background image

7 - 2

IP

N 07

4-

37

9-

P1

K

Cygnus Operating Manual

[1]

where:

D

1

= Initial density setting

T

x

= Thickness reading on Cygnus

T

m

= Measured thickness

7

A quick check of the calculated density may be made by programming the
instrument with the new density value and observing that the displayed
thickness is equal to the measured thickness, provided that the instrument's
thickness has not been zeroed between the test deposition and entering the
calculated density.

NOTE: Slight adjustment of density may be necessary in order to

achieve T

x

= T

m

.

7.3 Determining Tooling

1

Place a test substrate in the system's substrate holder.

2

Make a short deposition and determine actual thickness.

3

Calculate tooling from the relationship:

[2]

where

T

m

= Actual thickness at substrate holder

T

x

= Thickness reading on Cygnus

TF

i

= Initial tooling factor

4

Round off percent tooling to the nearest 0.1%.

5

When entering this new value for tooling into the program, T

m

will equal T

x

if

calculations are done properly.

NOTE: It is recommended that a minimum of three separate evaporations be

made when calibrating tooling. Variations in source distribution and other
system factors will contribute to slight thickness variations. An average
value tooling factor should be used for final calibration.