6 output name parameters, 1 query output names, 2 update output names – INFICON Cygnus Thin Film Deposition Controller User Manual

Page 127

Advertising
background image

5 - 23

IP

N 07

4-

37

9-

P1

K

Cygnus Operating Manual

5.4.3.6 Output Name Parameters
5.4.3.6.1 Query Output Names

Query: The message format is:

QO<Output Number>

Output Number = <byte> 0 to 38 (0 = All Outputs)

Command Response:

If Output Number was 1 to 38:

ACK<Output Name>|<Response Error Code>

If Output Number was 0:

ACK<Output 1 Name><Output 2 Name>...<Output 38 Name>|<Response
Error Code>

Output Name = <string> Up to 10 characters, null terminated.

5.4.3.6.2 Update Output Names

Update: The message format is:

UO<Output Number><Output Name>|<Output Number = 0><Output 1
Name><Output 2 Name>...<Output 38 Name>

Output Number = <byte> 0 to 38 (0 = All Outputs)

Output Name = <string> Up to 10 characters, null terminated

Command Response:

ACK|<Response Error Code>

NOTE: Output Names may only be updated if all channels are in Ready or Stop

and are not crystal switching on any channel.

Increment Counter x

Increment Counter x (x = 1 to 20)

0x4F

Start Deposit n

Start Deposit n

0x50

Goto NonDep Cnt n

Go to Non-Deposit Control n

0x51

Goto Post Dep n

Go to Post Deposit n

0x52

Continue Dep n

Continue Deposit n

0x53

RW Sample ON s

Leave RateWatch hold. (Can’t return
to hold until turned off.)

0x54

RW Sample OFF s

Allows return to RateWatch hold.

0x55

Table 5-7 Action Codes (continued)

Name

Description

Code