6 output name parameters, 1 query output names, 2 update output names – INFICON Cygnus Thin Film Deposition Controller User Manual
Page 127
5 - 23
IP
N 07
4-
37
9-
P1
K
Cygnus Operating Manual
5.4.3.6 Output Name Parameters
5.4.3.6.1 Query Output Names
Query: The message format is:
QO<Output Number>
Output Number = <byte> 0 to 38 (0 = All Outputs)
Command Response:
If Output Number was 1 to 38:
ACK<Output Name>|<Response Error Code>
If Output Number was 0:
ACK<Output 1 Name><Output 2 Name>...<Output 38 Name>|<Response
Error Code>
Output Name = <string> Up to 10 characters, null terminated.
5.4.3.6.2 Update Output Names
Update: The message format is:
UO<Output Number><Output Name>|<Output Number = 0><Output 1
Name><Output 2 Name>...<Output 38 Name>
Output Number = <byte> 0 to 38 (0 = All Outputs)
Output Name = <string> Up to 10 characters, null terminated
Command Response:
ACK|<Response Error Code>
NOTE: Output Names may only be updated if all channels are in Ready or Stop
and are not crystal switching on any channel.
Increment Counter x
Increment Counter x (x = 1 to 20)
0x4F
Start Deposit n
Start Deposit n
0x50
Goto NonDep Cnt n
Go to Non-Deposit Control n
0x51
Goto Post Dep n
Go to Post Deposit n
0x52
Continue Dep n
Continue Deposit n
0x53
RW Sample ON s
Leave RateWatch hold. (Can’t return
to hold until turned off.)
0x54
RW Sample OFF s
Allows return to RateWatch hold.
0x55
Table 5-7 Action Codes (continued)
Name
Description
Code