INFICON Cygnus Thin Film Deposition Controller User Manual

Page 119

Advertising
background image

5 - 15

IP

N 07

4-

37

9-

P1

K

Cygnus Operating Manual

47

(0x2f)

Dep after

PreDep

0 = Non Deposit

Control

1 = Deposit, No Auto

Soak

2 = Deposit, Auto Soak

Int

0

2

48

(0x30)

Rate Filter

Time

sec. 0.1, 0.4, 1.0, 4.0,

10.0, 20.0, 30.0

Float

0.1

30

49

0x31

Cross Talk

Percent

%

float

0.00

99.99

50

0x32

Auto-Z

1 = YES

0 = NO

Int

0

1

51

RateWatch

Option

0 = None

1 = Non Deposit

2 = Non Deposit and

Deposit

Int

0

2

NOTE 1: Outputs are numbered 1 through 38. Each can be used only once.
NOTE 2: Inputs are numbered 1 through 28. Each can be used only once.
NOTE 3: The number of Crucible Outputs required is dependent on the total number of crucibles. See

the following table. These outputs must be in consecutive order. The Crucible Output
parameter indicates the first output to be used. Therefore, the maximum value of this
parameter is dependent on the number of crucibles. For example, if the number of crucibles
is 16, and number of outputs is 4, and so the largest value acceptable for the crucible output
is 35, as outputs 35, 36, 37, and 38 will be used.

NOTE 4: The front panel display will show the time in Minutes:Seconds format, i.e., 5999 sec will be

shown as 99:59 mm:ss, etc.

Table 5-4 Channel Parameters (continued)

Channel

Command

ID

Name

Units / Allowed

Values / Notes

Data

Type

Low Limit

High Limit

Number of Crucibles

Number of Outputs

4

2

8

3

16

4

32

5

64

6