5 - 26

IP

N 07

4-

37

9-

P1

K

Cygnus Operating Manual

5

(0x05)

State

byte, encoding as follows:
0 = Ready
1 = Crucible Switch
2 = Rise 1
3 = Soak 1
4 = Rise 2
5 = Soak 2
6 = Deposit
7 = Time Power
8 = Non Deposit Control
9 = Non Deposit Hold
10 = Idle Ramp
11 = Manual
12 = Stop

6

(0x06)

State Time

Integer, seconds, time in the current state

7

(0x07)

Accumulated Time

Integer, seconds, time since Ready

8

(0x08)

Crystal Life

Integer, range will be 0 to 100

9

(0X09)

Crystals Remaining

Integer, range between 0 and 12, number of good
crystals on CrystalTwo, CrystalSix, Crystal12, or
Rotary.

10

(0X0A)

Crystal Position

byte, range between 1 and 12. The current crystal
position on a CrystalTwo, CrystalSix, or Crystal12

11

(0X0B)

Sensor State

byte, range between 0 and 2. 0 = Good crystal,
1 = Failed crystal, 2 = Invalid measurement on
crystal

12

(0x0C)

Fundamental
Frequency

Integer, unconverted. See Note 2.

13

(0x0D)

Fundamental Activity

Integer, range 0 to 650

Table 5-8 Status Channel Commands (continued)