INFICON Cygnus Thin Film Deposition Controller User Manual
Page 63

2 - 27
IP
N 07
4-
37
9-
P1
K
Cygnus Operating Manual
.
NOTE: If the crucible output type were normally closed (NC),
Table 2-2
would
need to be modified by exchanging open and closed.
4
Determine whether Turret Feedback is desired. This allows the turret position
controller to stop further instrument processing until the requested turret
position is satisfied. If chosen, a turret input must be connected to the turret
position controller’s feedback signal within one minute, else a STOP Crucible
ERR occurs.
If Turret Feedback is not chosen, program a Turret Delay Time, which allows
an adequate time for positioning to take place. Once the delay time has
expired, instrument state processing continues.
5
The selection of a particular crucible for a channel is defined in the CHANNEL
PROCESS display.
Table 2-2 Wiring To The Controller
Crucible
Position
Contact Status
Relay #6
Relay #7
1
Open
Open
2
Closed
Open
3
Open
Closed
4
Closed
Closed