1 status channel command – INFICON Cygnus Thin Film Deposition Controller User Manual
Page 129
5 - 25
IP
N 07
4-
37
9-
P1
K
Cygnus Operating Manual
5.4.4.1 Status Channel Command
The message format is:
SC<Command ID><Channel Number>
Command ID = <byte> See
Table 5-8
.
Channel Number = <byte> 0 to 6, 0 means all channels.
Command Response:
If Channel Number = 1 to 6:
ACK<<byte>|<Integer>|<float>|<Special>>|<Response Error Code>
See the "Response" column in
Table 5-8
.
If Channel Number = 0:
ACK<<byte><byte><byte><byte><byte><byte>|<integer><integer><integer>
<integer><integer><integer>|<float><float><float><float><float><float>|<Spec
ial><Special><Special><Special><Special><Special>>|<Response Error
Code>
Description: If Channel Number = 0, the response for each of the six channels
are given, channel 1 is given first.
Table 5-8 Status Channel Commands
SC
Command
ID (HEX)
Meaning
Response
0
(0x00)
Data
Byte Sensor Status, Byte State, Float Rate Filtered,
Float Thickness, Float Power, Float Rate Deviation
1
(0x01)
Filtered Rate
Float, A/s
2
(0x02)
Displayed Rate
Float, A/s, averaged over one second.
3
(0x03)
Source Power Output
Float, %
4
(0x04)
Thickness
Float, kЕ