Cygnus Operating Manual

UNABLE TO AUTO Z #

Indicates the requested Auto Z measurement cannot be completed due to the
crystal condition. This is caused by a weak or unstable first anharmonic
measurement. This also may be caused by the instrument not recognizing the
crystal as a new crystal, or inserting a used crystal.

VALUE TOO LARGE

Parameter entry is out of range. Allowable values will vary according to your
unit configuration or the parameter being defined. Press CLEAR to delete the
value and enter again.

VALUE TOO SMALL

Parameter entry is out of range. Allowable values will vary according to your
unit configuration or the parameter being defined. Press CLEAR to delete the
value and enter again.

XIU TEST PASSED

The XIU has passed the XIU test.

XIU TEST FAILED

The XIU has failed the XIU test.

XTAL FAIL

Indicates the crystal for the Channel is failed.

XTAL SWITCH ERROR

This message indicates a CrystalSix or Crystal12 sensor switcher failure. The
instrument did not detect the proper position after having attempted to rotate
the sensor head.

XTAL SWITCHING

Indicates a crystal switch is in progress.