4 isolated +24 v(dc) supply, Caution – INFICON Cygnus Thin Film Deposition Controller User Manual

Page 156

Advertising
background image

6 - 14

IP

N 07

4-

37

9-

P1

K

Cygnus Operating Manual

6.3.4.4 Isolated +24 V(dc) Supply

An isolated +24 V(dc) power supply is available on a 9-pin D-Sub connector on the
Cygnus back panel. This supply is rated for 1.75 Amps maximum.

The pin assignments for this connector are:

CAUTION

Both the isolated 24 V(dc) supply and the RS-232C
remote communications port use a 9-pin D-sub
connector. Care must be taken not to inadvertently
connect the RS-232C remote communications cable to
the 24-volt supply connector. Also, care must be taken
not to inadvertently connect the 24 V(dc) supply cable to
the RS-232C remote communications connector.

Table 6-5 +24 V(dc) Pin Connections

Pin

Function

1

Return

2

Return

3

Return

4

Not Connected

5

Not Connected

6

+24 Volts

7

+24 Volts

8

+24 Volts

9

Not Connected

Advertising