Cygnus Operating Manual

RATEWATCHER OPTION . . . . . . . . 0, 1, 2 (default = 0)

This parameter controls the use of the RateWatcher.
0 = None, disabled
1 = Non-deposit state only
2 = Non-deposit and deposit states

RATEWATCH HOLD TIME . . . . . . . 00:01 to 99:59 min:sec (default = 00:05)

RateWatch Time determines the time interval over which the sensor shutter will
be closed between samples.

RATEWATCH ACCURACY . . . . . . . 1 to 99 % (default = 5)

During the rate sampling period, the deposition rate is measured by the crystal;
source power control is active. When the rate is within the desired accuracy for
a period of time, the shutter is closed and the deposition state returns to HOLD.

RATE FILTER TIME. . . . . . . . . . . . . 0.1, 0.4, 1.0, 4.0, 10.0, 20.0, 30.0 sec.

(default = 0.1)

This parameter is used to apply a boxcar filter to the measured “raw” rate which
is used to produce the Rate DAC output voltage.
The Rate Filter Time may be used for enhanced resolution. If a source has a
long time constant, rate may be taken as constant over longer intervals
Averaging provides an opportunity to reduce rate noise. Depending on the
source of the noise, averaging can improve rate resolution:

Parameters

Averaging

Time

Rate

Filter
Time

RateWatcher

"Sample"

Acceptance

Window

Time

Rate and Power are simultaneously averaged

using the following stored data

0 min

< 10 sec

5 seconds

2 second interval before

most recent 0.5 seconds

0 min



10 sec

10 seconds

2 second interval before

most recent 0.5 seconds

1 min

any

60 seconds

2 second interval before

most recent 0.5 seconds

> 1 min

any

Averaging

Time * 60

seconds

Averaging Time excluding the most recently

calculated 1 minute average