3 parameter commands, 1 all parameters, 1 query all parameters – INFICON Cygnus Thin Film Deposition Controller User Manual

Page 113

Advertising
background image

5 - 9

IP

N 07

4-

37

9-

P1

K

Cygnus Operating Manual

5.4.3 Parameter Commands

Query commands are used to request established parameter values. There is a
specific query command for each parameter group. Depending on the actual
hardware configuration, the response to some Query commands may be an "E"
error code indicating there is no data to retrieve. Each command has one to three
arguments.

Update commands are used to change a specified parameter to a new value or
condition. Some parameters are order dependent in that they cannot be updated
prior to certain other parameters. There is a specific update command for each
parameter group.

The command variations include:

5.4.3.1 All Parameters

5.4.3.1.1 Query All Parameters

Query: The message format is:

QA

Command Response:

ACK<Returns values for the entire instrument configuration in the following order:

Length = 2 bytes (low byte / high byte)
All Channel 1 parameters, All Channel 2 parameters,...All Channel 6
parameters.
All General Parameters
All Logic Statements
All Output Types
All Output Names
All Input Names>|<Response Error Code>

NOTE: This length is NOT the same as the length of the packet. It tells how

many bytes follow in the data.

Table 5-3 Parameter Commands

QUERY command

parameter

UPDATE Command

QA

All Parameters

UA

QC

Channel Parameters

UC

QG

General Parameters

UG

QI

Input Name

UI

QL

Logic Statements

UL

QO

Output Name

UO

QT

Output Type

UT