3 parameter commands, 1 all parameters, 1 query all parameters – INFICON Cygnus Thin Film Deposition Controller User Manual
Page 113
5 - 9
IP
N 07
4-
37
9-
P1
K
Cygnus Operating Manual
5.4.3 Parameter Commands
Query commands are used to request established parameter values. There is a
specific query command for each parameter group. Depending on the actual
hardware configuration, the response to some Query commands may be an "E"
error code indicating there is no data to retrieve. Each command has one to three
arguments.
Update commands are used to change a specified parameter to a new value or
condition. Some parameters are order dependent in that they cannot be updated
prior to certain other parameters. There is a specific update command for each
parameter group.
The command variations include:
5.4.3.1 All Parameters
5.4.3.1.1 Query All Parameters
Query: The message format is:
QA
Command Response:
ACK<Returns values for the entire instrument configuration in the following order:
Length = 2 bytes (low byte / high byte)
All Channel 1 parameters, All Channel 2 parameters,...All Channel 6
parameters.
All General Parameters
All Logic Statements
All Output Types
All Output Names
All Input Names>|<Response Error Code>
NOTE: This length is NOT the same as the length of the packet. It tells how
many bytes follow in the data.
Table 5-3 Parameter Commands
QUERY command
parameter
UPDATE Command
QA
All Parameters
UA
QC
Channel Parameters
UC
QG
General Parameters
UG
QI
Input Name
UI
QL
Logic Statements
UL
QO
Output Name
UO
QT
Output Type
UT