11.6 Sample racks

■■■■■■■■■■■■■■■■■■■■■■

94

■■■■■■■■

915 KF Ti-Touch

Positions
Number of positions on the rack.

Code
The rack code corresponds to the alignment of magnets on the under-
side of the rack and is read out by the Sample Processor for rack recog-
nition.

[Load]

Load a new sample rack (see "Loading the sample rack", page 94).

[Copy]

Create a new sample rack by copying an existing rack (see "Creating a
new sample rack", page 95)
.

[Delete]

Delete the selected sample rack from the list.

[Edit]

Edit the data of the selected sample rack (see Chapter 11.6.1, page 95).

Loading the sample rack

When you receive the file for a new sample rack from your Metrohm rep-
resentative, you can easily import this file into your existing system. Pro-
ceed as follows:

1 Copy file to an external storage medium

Copy the file to a sub-directory of "Files".

If this structure is not maintained, the new rack will not be found
because the software directly accesses subdirectories of "Files".

Plug in the external storage medium at the Ti-Touch.

2 Display the list of saved sample racks

In the dialog Edit device / Sample racks, tap on the button
[Load].
The selection of file groups on the external storage medium is dis-
played. If only one group is available, then the list of the saved
sample rack files will be displayed directly.

Select the group with the desired sample rack.

Advertising