8 Titrants

915 KF Ti-Touch

■■■■■■■■

49

Input range

–999999999 - 9999999999

Default value

1.000

Unit of the titer.

Entry

10 characters maximum

Default value

empty

Selection

µmol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |
µg/L | ppm | % | mEq/L

NOTE

If you modify the titer of the titrant in a loaded determination at a later
date and would then like to recalculate the determination with the cor-
rected value, then you must modify the value in the determination data
under View data / Titrant data (see "Titrant data", page 201).

Date titer det.

Date and time of the last titer determination. For new titrants, the time
the preparation was made is specified until after the first time a titer deter-
mination has been carried out.

[Working life]

Define the working life of the sensor (see Chapter 8.3, page 50).

[Dosing unit]

This button is only displayed for Type = IDU.

Define the properties of the dosing unit used (see Chapter 8.4, page
51)
.

[Exchange unit]

This button is only displayed with Type = IEU or EU.

Define the properties of the exchange unit used (see Chapter 8.5, page
57)
.

[Titer options]

Display the properties for the titer determination (see Chapter 8.7, page
63)
.