28.1 Volumetric Karl Fischer titrations (KFT)

■■■■■■■■■■■■■■■■■■■■■■

266

■■■■■■■■

915 KF Ti-Touch

Selection

off

Default value

off

Filling rate

Rate at which the dosing cylinder is filled.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum filling rate depends on the cylinder volume (see Chapter
31, page 338)
.

28.1.5

Conditioning

The conditions required for conditioning are defined under [Condition-
ing]
.

Conditioning

on | off (Default value: on)

If this parameter is activated, then the first time the titration is started the
working medium will be titrated to the endpoint with the specified control
parameters. The status is kept stable. The actual method run does not
begin until [

] has been pressed once more.

Start drift

Conditioning OK will be displayed as soon as this drift has been reached
and the titration can be started.

Input range

1 - 999 µL/min

Default value

20 µL/min

Drift correction

The endpoint volume can be corrected by taking the drift value into
account. For this, the drift is multiplied with the drift correction time and
this value is then subtracted from the endpoint volume. The drift correc-
tion time is the time interval between the end of conditioning and the end
of the determination.

Selection

auto | manual | off

Default value

off

auto
The value of the current drift is automatically applied at the start of the
titration.
manual
If the drift is known throughout a longer period of time, this can be
entered manually.