915 KF Ti-Touch

■■■■■■■■

161

Method check at start

on | off (Default value: on)

If this option is activated, then the following tests are carried out at the
method start.

Are the required control and peripheral devices connected?

Are the titrant and the sensor on hand?

Have the monitoring intervals expired?

etc.

NOTE

We recommend that the method check only be deactivated if the titra-
tion must begin immediately after the start of the determination (e.g.
with very rapid reactions such as enzyme-catalyzed reactions). If the
method check is deactivated, it could happen that the determination
run will be interrupted, e.g. because devices, titrants or sensors are not
on hand.

The following settings can be implemented for a connected USB Sample
Processor:

Rack reset

on | off (Default value: off)

If this option is activated, then the sample rack is initialized at the method
start. The following actions are carried out:

The lift is moved upwards.

The sample rack is rotated to the starting position.

The rack code of the rack attached is being read out and the corre-
sponding rack data is transferred to the Sample Processor.

The sample variable is being reset to the value 1.

Check rack

Definition of the sample rack which must be attached at the time of the
method start. This ensures that the method will be carried out only with
this sample rack.

NOTE

The sample rack can however only be checked if theRack reset option
is activated.