28 Parameters

915 KF Ti-Touch

■■■■■■■■

315

Input signal

Selection of the signal out of the templates or entering the required bit
pattern. Templates are defined under System

Templates Input

lines.

Entering a bit pattern:

0 = line inactive

1 = line active

* = retain line status

The input lines are always numbered from right to left, i.e. with the signal
*******1 line 0 is expected to be active.

NOTE

We recommend masking lines that are of no interest or for which no
defined condition can be predicted with an asterisk (*).

Entry

Bit patterns containing exactly 8 characters
or a max. of 24 characters for the name of
the template

Default value

********

Selection

Selection of the templates defined

Timeout

When this time interval has expired without recognizing the remote signal,
the action defined in the following is triggered.

Input range

0 - 999 s

Default value

0 s

Action

Selection of the action which is carried out when the time interval has
expired.

Selection

Display message | Document message | Cancel
determination

Default value

Display message
For all three options it is documented in the deter-
mination data (see dialog More determination
data / Messages
), that the time interval has been
expired.

Display message
A message is displayed. You can select whether you want to continue
with the determination or cancel the run.