31.6 Result variables as parameter setting

■■■■■■■■■■■■■■■■■■■■■■

344

■■■■■■■■

915 KF Ti-Touch

31.6

Result variables as parameter setting

For most of the numerical parameters, the result of a calculation can also
be utilized in place of a number. This result must be calculated in a pre-
ceding calculation command (command CALC). A typical application is
the use of a relative start volume.

The following example is used to explain how this method is used to
define a sample size-dependent start volume for a titration. Proceed as
follows:

1 Create a new calculation

Insert the calculation command CALC in front of the titration
command.

Enter the calculation formula, e.g. R1 = C00*3.

2 Define a result variable as parameter setting

In the titration command, enter the variable R1 as the start vol-
ume.