31 Appendix

915 KF Ti-Touch

■■■■■■■■

339

31.2

Exchange unit

31.2.1

Maximum dosing and filling rate

The maximum dosing rate and maximum filling rate for the exchange unit
depend on the cylinder volume:

Cylinder volume

Maximum rate

1 mL

3.00 mL/min

5 mL

15.00 mL/min

10 mL

30.00 mL/min

20 mL

60.00 mL/min

50 mL

150.00 mL/min

Independent of the cylinder volume, values ranging from 0.01 to 166.00
mL/min can always be entered. When the function is carried out the rate
will be, if necessary, decreased automatically to the highest possible value.

31.2.2

Default parameters for preparing (PREP)

The PREP command is used to rinse and fill the cylinder and tubings of
the exchange unit air bubble-free. You should carry out this function
before the first determination or once per day.

If no titrant is selected in the command, preparing will be carried out with
the following default parameters:

The entire cylinder volume is dosed twice at the maximum dosing rate.

The cylinder is filled at the maximum filling rate.

31.3

Stirring rate

The stirring rate can be adjusted in steps of –15 to +15.

The approximate rotational speed can be calculated with the following
formula:

Rotational speed/min (r/min) = 125 · stirring rate

Example:

Stirring rate set: 8

Rotational speed in rpm = 125 · 8 = 1000