3 variable list – Metrohm 915 KF Ti-Touch User Manual

Page 310

Advertising
background image

28.4 Calculations

■■■■■■■■■■■■■■■■■■■■■■

296

■■■■■■■■

915 KF Ti-Touch

28.4.3.3

Variable list

The following table contains all of the variables which can be used for cal-
culations. For variables having an index (e.g. 'EP1'), the index must be
entered manually. In the following table, this index is characterized with
the character "#".

Variable

Description

C00

Sample size

CI1, CI2

Sample identifications

The sample identifications can be used in calculations only if numerical values are entered.

DD

Duration of the entire determination

Titrants

TITER

Titer of the titrant selected in the titration command

CONC

Concentration of the titrant selected in the titration command

Titrations, measurements, calibrations

EP#

Volume/quantity of endpoint EP# (# = 1…9)

EC#

Charge at endpoint EP# (# = 1…9)

EM#

Measured value of endpoint EP# (# = 1…9)

EF#

ERC of endpoint EP# (# = 1…9)

ET#

Temperature at endpoint EP# (# = 1…9)

ED#

Time at endpoint EP# (# = 1…9)

ESI#

Recognition of endpoint EP# (# = 1…9)

Endpoint found = 1, no endpoint = 0

MIM

Initial measured value, i.e. measured value prior to the processing of the start conditions

MIT

Initial temperature, i.e. temperature prior to the processing of the start conditions

MSA

Volume for start volume

MSP

Volume for start measured value

MSS

Volume for start slope

MSV

Volume for all the start conditions

MSD

Duration of start conditions

MSM

Start measured value, i.e. measured value after the processing of the start conditions

MST

Start temperature, i.e. temperature after the processing of the start conditions

MCD

Total duration of the titration, measurement or calibration

MTM

Type of temperature measurement (Pt1000, NTC or manually)

Format = Text

MDD

Duration of effective dosing, i.e. without filling times, pauses.