Metrohm 915 KF Ti-Touch User Manual

Page 266

Advertising
background image

27.6 USB Sample Processor

■■■■■■■■■■■■■■■■■■■■■■

252

■■■■■■■■

915 KF Ti-Touch

Input range

0 - 'maximum stroke path' mm
A lift position of 0 mm corresponds to the "home
position", i.e. the lift is located at the upper stop
position. The maximum stroke path is defined in the
properties of the tower (see "Maximum stroke
path", page 88)
. If a higher value is entered, an
error message will be displayed.

Selection

Work position | Shift position | Rinse position |
Special position

Default value

Work position

Lift rate

Rate at which the lift is moved in the manual control.

Input range

5 - 25 mm/s

Default value

25 mm/s

[Assign lift pos.]

Assign the current lift position to a preset lift position.

Assigning lift positions
Specific lift positions are stored separately for each sample rack used as
well as for Tower 1 and Tower 2. The following specific lift positions are
available:

General rack positions
A work position, a shift position, a rinse position and a special position
can be defined for general rack positions.

Special beaker positions
An individual specific work position can be defined for each special
beaker. The definitions of the general rack positions are used for the
shift, rinse and special positions of the affected tower.

External positions (only with robotic arm and mounted Swing Head,
see Chapter 27.6.3, page 254)
For each of the four possible external positions, a specific working
position can be defined. Shift and rinse positions can only be defined
for all four external positions together. A special position is not possi-
ble.

Proceed as follows to assign the current lift position to a specific lift posi-
tion:

1 Move to lift position

Enter the desired lift position in mm and tap on [Start].

The lift moves to the desired lift position.