2 pausing the method run (wait), Pausing the method run (wait) – Metrohm 915 KF Ti-Touch User Manual

Page 342

Advertising
background image

28.9 Miscellaneous commands

■■■■■■■■■■■■■■■■■■■■■■

328

■■■■■■■■

915 KF Ti-Touch

Input range

0 - 999 s (Increment: 1)
During this time the stirrer is running.

Selection

on | off

Default value

on

on
The stirrer is switched on. The stirrer runs until it is explicitly switched
off. If the determination is canceled with the key [

], the stirrer is

also switched off.
off
The stirrer is switched off.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 rpm. The formula for calculating the rota-
tional speed is specified in chapter 31.3, page 339. The optimum stirring
rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

–15 - 15

Default value

8

28.9.2

Pausing the method run (WAIT)

The method run can be paused with the command WAIT.

Hold sequence

on | off (Default value: off)

If this parameter is activated, then the method run is paused until it is con-
tinued manually.

Waiting time

This parameter can only be edited when Hold sequence is deactivated.

The method run is automatically continued after this waiting time.

Input range

0 - 99999 s

Default value

30 s

Message

on | off (Default value: off)

If this parameter is activated, the text message defined in the following is
displayed during the waiting time.

Advertising