11.6 Sample racks

■■■■■■■■■■■■■■■■■■■■■■

98

■■■■■■■■

915 KF Ti-Touch

Input range

0 - 'Maximum stroke path' mm

Default value

0 mm
A lift position of 0 mm corresponds to the "home
position", i.e. the lift is located at the upper stop
position. The maximum stroke path is defined in the
properties of the tower (see "Maximum stroke
path", page 88)
.

Shift position

The lift is raised to this position with each rotational movement of the rack
whenever it is located at a lower lift position.

Input range

0 - 'maximum stroke path' mm
A lift position of 0 mm corresponds to the "home
position", i.e. the lift is located at the upper stop
position. The maximum stroke path is defined in the
properties of the tower (see "Maximum stroke
path", page 88)
.

Special position

For sample positions, you can define an additional lift position for special
applications.

Input range

0 - 'maximum stroke path' mm
A lift position of 0 mm corresponds to the "home
position", i.e. the lift is located at the upper stop
position. The maximum stroke path is defined in the
properties of the tower (see "Maximum stroke
path", page 88)
.

Editing special beakers
Sample rack list: Rack

Edit Special beakers