Metrohm 915 KF Ti-Touch User Manual

Page 357

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

31 Appendix

915 KF Ti-Touch

■■■■■■■■

343

Assignment

Pin No.

Function

Output 13

20

0 volts / GND

14

+5 volts

15

0 volts / GND

25

Table 9

Explanation of the individual functions

Function

Explanation

Start

The current method is started at the time of acti-
vation.

t

pulse

> 100 ms

Stop

The current method is canceled (Stop) at the time
of activation.

t

pulse

> 100 ms

Quit

The current command in the determination run
will be canceled at the time of activation.

t

pulse

> 100 ms

Ready

The instrument is ready to receive a start signal.

Conditioning OK

The line is set when Conditioning with KFT titra-
tion is at OK. The line remains set until the deter-
mination is started with [START].

Determination

The instrument performs a data-generating
determination.

EOD

End of Determination.

Pulse (t

pulse

= 200 ms) after a determination.

Error

The line is set for error message display.

Warning

The line is set for warning message display.