8.5 Exchange unit

■■■■■■■■■■■■■■■■■■■■■■

58

■■■■■■■■

915 KF Ti-Touch

Entry

24 characters maximum

Default value

empty

Serial number

Serial number of the exchange or dosing unit. It will be read out automati-
cally on units with integrated data chips.

Entry

8 digits maximum

Cylinder volume

Cylinder volume of the exchange unit. It will be read out automatically on
exchange units with integrated data chips.

Selection

1 | 5 | 10 | 20 | 50

Default value

20

Serial no. cyl.

Serial number of the dosing cylinder. It will be read out automatically on
exchange or dosing units with integrated data chips. The number can be
changed manually at any time, e.g. when a cylinder is replaced.

Entry

8 digits maximum

[GLP test]

Define the time interval for the GLP test (see Chapter 8.6, page 61).

[PREP param.]

Enter the parameters for the preparation (see Chapter 8.5.1, page 58).

[Tubing param.]

Enter the parameters for the connected tubings (see Chapter 8.5.2, page
59)
.

8.5.1

Parameters for the preparation (PREP)

Titrant: Edit

Exchange unit PREP param.

In the dialog Exchange unit / PREP parameters, you can adjust the
parameters for the execution of the Prepare function (command PREP).
This function is used to rinse the cylinder and tubings of the exchange unit
and fill it air bubble-free. You should carry out this function before the
first determination or once per day.

Volume

Volume of titrant dosed during a rinsing cycle.

Input range

0.00000 - 99999.9 mL

Selection

Cylinder volume

Default value

Cylinder volume